Click to edit Master title style,*,Click to edit Master text styles,Second level,Third level,Fourth level,Fifth level,Internal use only,飞利浦医疗保健,zhang hua,Jun.29,2010,Click to edit Master title style,Click to edit Master text styles,Second level,Third level,Fourth level,Fifth level,*,Internal use only,飞利浦医疗保健,zhang hua,Jun.29,2010,Click to edit Master title style,Click to edit Master text styles,Second level,Third level,Fourth level,Fifth level,*,Internal use only,飞利浦医疗保健,zhang hua,Jun.29,2010,Click to edit Master title style,Click to edit Master text styles,Second level,Third level,Fourth level,Fifth level,*,飞利浦医疗保健,超声设备的维修和保养,飞利浦医疗保健超声设备的维修和保养,1,2,内容,电网和用电安全,地线,日常维护,IU22,和,IE33,飞利浦超声的简述,故障代码的分析和排除,2内容电网和用电安全,2,3,正确使用交流电,提议:超声不要跟电脑、电风扇等其他设备共用一个插线板。,零线 火线,反 接,使用廉价插线板,能插就插,插满为止,不妥的行为和习惯:,忽视接地,3正确使用交流电提议:超声不要跟电脑、电风扇等其他设备共用一,3,4,正确的接线,(,单项,220V,50Hz),左零、右火、中间地。,地线,黄绿花,火线,棕色,零线,蓝色,4正确的接线 (单项,220V,50Hz)左零,4,5,每个季度检查一次地线,请求助电工或飞利浦公司。,万用表是简易可行的方法,交流档检查零地电压,要求小于,2 V,。,ErrorId:0,File:D:Viewsusdbtlbca1ms5b5_g2b_rel_fec_svborisRoot borisHardwarefecappmainfecMonitor.cpp,Line number:1713,Info Msg,:Error in calibrateHVi().Quiescent Current HVI-=0 x1ae=26 mA,FEC ASSERT(Print Only)-Expression:0,Info Msg:Unable to get semaphore in CFecOutMsg(),*,Power Mon Calibration Failed,!*,提示前端有过流的问题,但无法确认是哪块,PCB,现场准备,最小配置隔离,AFP,,,NAIM,、,FEC,,可以,解决问题,33 IE33,开机报错20,33,34,Q&A,34 Q&A,34,临床工程师培训无锡超声设备的维修和保养课件,35,